4 level and usage queries, 5 length configuration commands and queries, Level and usage queries – American Magnetics 187 Self-Compensating Liquid Level Controller User Manual
Page 55: Length configuration commands and queries, Remote interface reference
44
Remote Interface Reference
Level and Usage Queries
W
ALARm:RATE?
Returns the RATE alarm value in the presently selected remote units per
minute.
W
CONFigure:ALARm:MUTE {0|1}
or
{NO|YES}
Mutes the continuous audible alarm if a HI, LO, or Over/Under Flow
alarm condition exists. Error conditions will still report as a short audible
beep. “0” or “NO” enables the continuous audible alarm. “1” or
“YES”disables (or mutes) the continuous audible alarm.
W
ALARm:MUTE?
Returns “1” if the continuous audible alarm is muted, “0” otherwise.
4.5.4
Level and Usage Queries
The level and usage queries return the current level and usage rate in the
presently selected remote units.
W
LEVel?
Returns the level in the presently selected remote units.
W
USAGe?
Returns the usage rate in the presently selected remote units per minute.
4.5.5
Length Configuration Commands and Queries
The length configuration commands and queries control the zero offset and
measured span settings for the Model 187. For more information regarding
the length configurations, see paragraph 3.1.3 on page 14. It is important
to limit the zero offset and measured span settings so as to only measure
the span between the reference sensors.
Attempting to set or query for the length settings in the percent remote
units will generate an error. The selected remote units must be inches or
centimeters.
W
CONFigure:LENgth:ZERO <
zero_offset
>
Configures the “zero point”, or the zero offset from the bottom of the sensor
which will function as the “zero level”. The argument must be in the
selected remote units.
W
LENgth:ZERO?
Returns the “zero point” or zero offset in the selected remote units.