SECTION 13—BATTERIES
Part No. 1143206
59
ƽ WARNING
Batteries with terminal configuration as shown below MUST be used. Batteries that
have the reverse terminal configuration MUST NOT be used - otherwise injury and
damage may occur.
NOTE: Recommended battery type is spill proof and requires no maintenance except routine
charging.
NOTE: Charge batteries daily. DO NOT allow batteries to completely discharge. To charge
batteries, refer to Charging Batteries on page 63.
NOTE: Invacare recommends that both batteries be replaced if one battery is defective.
Removing/Installing the Batteries
NOTE: For this procedure, refer to FIGURE 13.1 on page 60.
ƽ WARNING
Install protective caps on POSITIVE (+) and NEGATIVE (-) terminals.
All battery terminal covers (two on the right battery and two on the left battery)
MUST be installed prior to use.
Removing
1. Disconnect the joystick. Refer to Disconnecting/Connecting the Joystick on page 56.
2. Remove the seat. Refer to Removing/Installing the Seat Assembly on page 37.
3. Remove the top shroud. Refer to Removing/Installing the Top Shroud on page 49.
4. Disconnect the right battery from the controller (BLACK connector).
DO NOT USE
NEGATIVE (-)
Battery Terminal
POSITIVE (+)
Battery Terminal
U1 Battery
U1 Battery
POSITIVE (+) Battery Terminal
NEGATIVE (-)
Battery Terminal
USE THIS CONFIGURATION