Activeforever Mettler Sonicator 740X Therapeutic Ultrasound Device User Manual
Sonicator, 740 specifications, General specifications
Sonicator
®
740 Specifications
General Specifications:
ETL and C-ETL Listed:
Model ME 740 (9801427)
Input:
100-240 VAC ±10%
Frequency:
60Hz
±3 and 50Hz ±3
Input Current:
less than 0.6 amps
Optional
Battery:
Rechargeable Smart Lithium Ion Battery Pack rated at 10.8Vand 4.8Ah
Patents:
U.S. and foreign patents applied for and granted including U.S. Patent
Numbers
4,966,131
and
5,095,890
Timer Accuracy:
±10% of the indicated time
Maximum treatment time:
29 minutes
Size:
13.5 in (L) x 9.5 in (W) x 7 in (H)
Weight:
3.75
pounds
without
battery
4.5
pounds
with
battery
Ultrasonic Generator Specifications:
Frequency:
1.0 MHz ±10%
3.2
MHz
±10%
3.3
MHz
±10%
Modes:
Continuous
Pulsed 10, 20 and 50% Duty cycle
Pulse repetition rate:
100 Hz ±5% (Pulse Mode)
Pulse duration:
1, 2 or 5 msec ±5%
Temporal peak/average
intensity ratio:
10:1, 5:1 and 2:1 ±5%
Maximum output power:
22 W with a 10 cm² applicator, (ME 7410)
11 W with 5 cm² applicator, (ME 7413)
2.2 W with a 1 cm² applicator (ME 7431)
Maximum intensity:
2.2 W/cm²
Indication accuracy:
±20% (for any level above 10% of maximum)
Ultrasonic Applicator Specifications:
Piezoelectric discs:
The output transducer utilizes a barium titanate disc with a specially
coated face.
Individual Applicator Specifications:
Applicator Part Number
Frequency
ERA
ME 7513
1 and 3.2 MHz ±10%
5 cm² ±20%
ME 7310
1 MHz ±10%
10 cm² ±20%
ME 7331
3.3 MHz ±10%
1 cm² ±20%
Maximum Beam
Uniformity
Ratio:
6:1
1333 South Claudina Street • Anaheim, California 92805 USA • Tel: 1 (714) 533–2221 • FAX: 1 (714) 635–7539