Color Measurement Instruments

Calibrating the DTP41

For the best color accuracy, calibrate the X-Rite DTP41 spectrophotometer every time
you calibrate the EX7750. You can initiate the calibration sequence in ColorWise Pro
Tools Calibrator as part of the process of EX7750 calibration. The computer may also
prompt you that it is necessary to calibrate the spectrophotometer.

Alternatively, you can initiate the DTP41 calibration process using the Instrument
button on the DTP41. For more information about this method, see the
documentation that accompanies the X-Rite DTP41.

Multiple users can be connected to one EX7750 with ColorWise Pro Tools, but only
one user at a time can use the Calibrator module. An error message appears if you try
to calibrate when another user is already using ColorWise Pro Tools to calibrate.

You will need the Color Reflection Reference strip that is included with the
spectrophotometer. Remove the strip from its protective envelope, handling it by the
edges only. Keep the strip free of dust and smudges and always store it in the protective
envelope.

T

O

CALIBRATE

THE

X-R

ITE

DTP41

1.

Connect the spectrophotometer to the computer and turn on power (see

page A-3

).

2.

Start Calibrator.

For instructions on starting Calibrator, see

page 2-6

.

3.

Choose DTP41 as the measurement method.