Pulse modulation rise and fall time, Modulation oscillator – VXI 3002 User Manual
Page 95

5-20
Pulse modulation rise and fall time
C2567
OUTPUT
Function generator
RF
OUT
UUT
PULSE
INPUT
Oscilloscope
Fig. 5-12 Pulse modulation rise and fall time test set-up
Test procedure
(1) Connect the test equipment as shown in Fig. 5-12.
(2) On the UUT set:
Carr Freq
50 MHz
RF Level
+7 dBm
Pulse
ON
(3) Set the function generator to produce 10 kHz, 0 V to +5 V square wave.
(4) Adjust the oscilloscope controls such that the rise time of the envelope can be measured.
(5) Measure the rise time between the 10% to 90% points checking that it is within the
specification shown in Table 5-42.
(6) Repeat (4) to (5) for the fall time of the envelope.
Modulation oscillator
Specification
Frequency range
0.01 Hz to 20 kHz.
Resolution
0.01 Hz for frequencies up to 100 Hz,
0.1 Hz for frequencies up to 1 kHz,
1 Hz for frequencies up to 20 kHz.
Frequency accuracy
As frequency standard.
Distortion
Less than 0.1% THD at 1 kHz.
Waveforms
Sine to 20 kHz, triangle or square wave to 3 kHz.
Audio output
The modulation oscillator signal is available on a front-panel BNC connector at a level of
2 V RMS EMF from a 600
Ω source impedance.
Test equipment
Description Minimum
specification Example
Frequency
counter
9 kHz to 2.4 GHz
Agilent 53181A
with option 030
Audio analyzer
Capable of measuring THD of 0.01% at 1 kHz
Rohde &
Schwarz UPA3