Am scale shape – VXI 3002 User Manual
Page 91

5-16
(8) Set the UUT to RF level +19 dBm and repeat (2) to (5) using Table 5-33.
AM scale shape
Test procedure
(1) Connect the test equipment as shown in Fig. 5-7.
(2) On the UUT set:
Carr Freq
100 MHz
RF Level
0 dBm
Mod Mode
AM Internal
Source
On
Mod
On
AM1
Level
30%
(3) On the modulation meter, select CAL, AM, 300 Hz
Þ
3.4 kHz LF filter.
(4) Measure the AM accuracy at the depths shown in Table 5-34 checking that the results are
within specification.
External AM frequency response (ALC off, DC coupled)
Test procedure
100 Hz to 30 kHz
(1) Connect the test equipment as shown in Fig. 5-8.
(2) On the UUT set:
Carr Freq
400 MHz
RF
Level
−4 dBm
Mod Mode
AM External
AM1
Level
80%
Source
On
Mod
On
DC
coupling
(3) Set the function generator to give 1 V RMS, 1 kHz sine wave.
(4) On the modulation meter select CAL, AM, 10 Hz
Þ
300 kHz filter.
(5) On the modulation meter check that the AM reading is between 76% and 84%, then set a
reference using the relative function.
(6) Record the absolute reading for use in the formula in (16) below.
(7) Set the function generator to each of the frequencies shown in Table 5-35 checking that the
relative readings on the modulation meter are within specification.
(8) Set the UUT RF level to +7 dBm and repeat (3) to (7) using Table 5-36.
(9) Set the UUT RF level to +13 dBm and repeat (3) to (7) using Table 5-37.