beautypg.com

Default settings – VXI 3002 User Manual

Page 35

background image

3-5

Default settings

The instrument is reset to the factory default settings in the following cases:

(1) At power-up.

(2) Following execution of the RCL 999 command.

(3) Following execution of the *RST command.

The default settings are shown in Table 3-1 .

Table 3-1 Instrument default settings

Carrier frequency

:

Step

:

2.4 GHz
1 kHz

RF level
Step

:

Status:

RF output :

−137 dBm

1 dB

ON
Enabled

Modulation mode

Internal FM, modulation disabled

Modulations

:

FM1

:

Deviation: 0 Hz, OFF

:

Internal source, frequency: 1 kHz, sine

FM2

:

Deviation: 0 Hz, OFF

:

Internal source, frequency: 400 Hz, sine

ΦM1

:

Deviation: 0 rad, OFF

:

Internal source, frequency: 1 kHz, sine

ΦM2

:

Deviation: 0 rad, OFF

:

Internal source, frequency: 400 Hz, sine

AM1

:

Deviation: 0%, OFF

:

Internal source, frequency: 1 kHz, sine

AM2

:

Deviation: 0%, OFF

:

Internal source, frequency: 400 Hz, sine

Pulse

:

OFF

Modulation steps :

∆FM 1 kHz, ∆ΦM 0.1 rad, ∆AM 1%

Mod frequency steps :

10 Hz

Carrier sweep
Freq

mode:

Mode

:

Type:

Ext trigger :

Start

:

Stop:
Step

size:

Time:


Fixed
Single sweep
Linear
OFF
9 kHz
2.4 GHz
1 kHz
50 ms