Pulse modulation on/off ratio – VXI 3002 User Manual

Page 94

background image

5-19

(7) Set the UUT RF level to +14 dBm and repeat (5) using Table 5-40.

Pulse modulation on/off ratio

C5358

OUTPUT

Function generator

RF

OUT

UUT

PULSE

INPUT

RF INPUT

Spectrum analyzer

Fig. 5-11 Pulse modulation on/off ratio test set-up

Test procedure

(1) Press CAL on the spectrum analyzer.

(2) Connect the test equipment as shown in Fig. 5-11.

(3) On the UUT set:

Carr Freq

32 MHz

RF Level

0 dBm

Pulse

ON

(4) Set the function generator to provide +5 V DC. The RF output will now be enabled

(5) Tune the spectrum analyzer to the same frequency as the signal generator.

(6) Press PEAK FIND on the spectrum analyzer and note the output level.

(7) Apply a short circuit to the PULSE INPUT socket.

(8) Again note the output level measured by the spectrum analyzer.

(9) The difference between the levels recorded in (6) and (8) is the pulse mod on/off ratio.

Check that the ratio is within specification using Table 5-41.

(10) Repeat (5) to (9) for each of the frequencies shown in Table 5-41.