Verifying imported rules, Dicom gateway configuration – Vista Imaging Vista Routing User Manual
Page 53

Routing User Guide
Configuring Routing
4
Open a terminal window and log into the Routing Gateway software.
5
Enter
3
to select “Routing Gateway” from the DICOM Gateway Menu [
MAGDMENU
].
6
Enter
5
to select “Import Routing Rules.”
7
When you are prompted to build the Routing Table, enter
Y
.
Ready to build the Routing Table? y// y
8
The file will be imported. If the import is successful, the following will be displayed.
You will be prompted to return to the menu.
nnn rules stored in the database
Press
 If you encounter any errors, record the information and then press Enter. You will
need to correct the offending rules and repeat these steps.
Verifying Imported Rules
When new routing rules are imported it is important to double-check that the new set of
rules corresponds to the intent of the end-user.
To verify imported rules
Open a terminal window and log into the Routing Gateway software.
1
Enter
3
to select “Routing Gateway” from the DICOM Gateway Menu [
MAGDMENU
].
2
Enter
10
to select “Display Routing Rules”.
3
The current set of Routing Rules will be displayed:
April 2006
VistA Imaging V. 3.0, Patch 18
45