Page 93

Using the Menus
1760 Series Component Waveform/Vector Monitor User Manual
3-25
and
RESET VCAL
cancels any vertical gain adjust-
ments; the instrument returns to the factory cali-
brated gain settings.
H In vector mode, the center bezel control adjusts the
vector signal gain. Once gain has been adjusted,
RESET VCAL
appears on screen. Selecting
RESET
VCAL
cancels any vector gain adjustments and
returns the instrument to the factory calibrated gain
settings.
REMOTE
H The currently operating software version is dis-
played in the lower right corner of the screen.
H Three menu choices are available.
REM INPUT
configures rear-panel
REMOTE
connector pin 3 to
accept External Horizontal or Staircase input.
STAIRCASE
configures the staircase input for 3-step
(RGB)
or 4-step
(YRGB)
signal.
BAUD
configures the
data transmission rate for the rear-panel RS232
connector for 1200, 4800, or 9600 baud.
H The remote input is enabled by a TTL low or switch
to ground. See page 3-29.
REFERENCE
(1765 Only)
H NTSC external reference and PAL external refer-
ence can be set to
EXT, A, A1, A2, A3, B, B1, B2,
or
B3
.
H When The front-panel
REF EXT
is selected (LED
lights) the instrument uses the rear-panel
EXT REF
input. Dual-standard monitors can be configured for
two external references; one for NTSC and one for
PAL. Any of the input channels can be designated
as a second external reference by selecting that input
channel as the
NTSC EXT REF
or
PAL EXT REF
through this menu.