Table a- 17: environmental characteristics, Table a- 18: physical characteristics – Tektronix 1760 User Manual
Page 142

Appendix A: Performance Specification
A-14
1760 Series Component Waveform/Vector Monitor User Manual
Table A- 17: Environmental Characteristics
CATEGORY
DESCRIPTION
Operating Temperature
Req:
0
° to 50° C (+32° to 122° F)
Storage Temperature
Req:
--40
° to 75° C (--40° to 158° F)
Operating Altitude
Req:
To 15,000 feet (4572 meters)
Storage Altitude
Req:
To 50,000 feet (15,240 meters)
Vibration
Req:
5 minutes at 5 -- 15 Hz with 0.060 inch dis-
placement
5 minutes at 15 -- 25 Hz with 0.040 inch dis-
placement
5 minutes at 25 -- 55 Hz with 0.020 inch dis-
placement
Military Specification: Mil--T--28800D,
Paragraph 1.2.2, Class 3
Mechanical Shock
Req:
Non Operating: 50 g’s 1/2 sine, 11 ms duration
3 shocks per surface (18 total)
Transportation
Req:
Qualified under NSTA Test Procedure 1A,
Category II (24 inch drop)
Humidity
Req:
Will operate at 95% relative humidity for up to
five days. Do not operate with visible moisture
on the circuit boards.
Table A- 18: Physical Characteristics
CATEGORY
DESCRIPTION
Dimensions
Req:
Height:
5 1/4 inches (133.4 millimeters)
Width:
8 1/2 inches (215.9 millimeters)
Depth:
18 1/8 inches (460.4 millimeters)
Weight
Req:
Net: 8 pounds (3.8 kilograms)
Shipping: Approximately 15.7 pounds (7.2
kilograms)