Tektronix 1760 User Manual
Page 126

Calibration
4-26
1760 Series Component Waveform/Vector Monitor User Manual
Set Horizontal Gain
1.
To set horizontal gain, push the
CONFIG
menu button.
Select
CALIBRATE
, then
CAL SIG ON
and
100 KHZ
.
Select
GAIN CAL ON
.
a.
In
1 LINE SWEEP
(5
µs/div), adjust the
HORIZ CAL
control for one-half cycle per major graticule
division; in
2 LINE SWEEP
(10
µs/div), adjust for one
full cycle per major division.
b.
Pushing
RESET H CAL
will cancel any gain adjust-
ment and reinstate the factory gain setting.
Set Vector Gain
1.
To adjust vector gain, select
VECTOR
display mode and
push the
CONFIG
menu button. Select
CALIBRATE
, then
CAL SIG ON
and
GAIN CAL ON
.
a.
Adjust
VECTOR CAL
so that the calibration signal
circle falls on the compass rose (outer ring).
b.
Pushing
RESET V CAL
will cancel any gain adjust-
ment and reinstate the factory gain setting.