Table a- 3: waveform horizontal deflection – Tektronix 1760 User Manual
Page 134
Appendix A: Performance Specification
A-6
1760 Series Component Waveform/Vector Monitor User Manual
Table A- 3: Waveform Horizontal Deflection
CATEGORY
DESCRIPTION
Sweep
Req:
Synchronization: Sweep triggered by horizon-
tal and vertical sync pulses
RI:
Sweep Length:
≈12 divisions
RI:
Sweep freeruns without input
Sweep Timing Accuracy
Req:
1 Line: 5 s/division
±1%
2 Line: 10 s/division
±1%
RI:
1 Field displays one full field,
including field rate sync.
2 Field displays two full fields and
the field rate sync between them.
Sweep Linearity
Req:
1 Line:
±1%
2 Line:
±1%
Magnified Sweep Accuracy
Req:
1 Line: 0.2 s/division
±1%
2 Line: 1.0 s/division
±1%
Magnified Sweep Linearity
Req:
1 Line:
±1%
2 Line:
±1%
Horizontal Position Range
Req:
Any portion of the synchronized sweep can be
positioned on screen in all sweep modes.
External Horizontal Input
Req:
2 divisions/volt,
±2%
RI:
Menu is selected and enabled by RE-
MOTE connector ground closure.
Remote Sync
RI:
Input Amplitude: TTL level
RI:
Frequency: 25 Hz to 100 Hz positive
edge-triggered sweep
RI:
Enabling Signal: TTL low or ground
closure