Important, Precautions when unpacking the microscope – Olympus IX81 User Manual

Page 6

background image

1

IMPORTANT

The IX81 microscope can be used in stand-alone mode (operated from the U-HSTR2 hand switch) or can operated from a
PC when the IX2-BSW software (Ver. 01.03, compatible with Windows 2000/Me) is installed in the PC.

Be sure to prepare the U-ZPCB (T2) Z-board and attach it to the IX81, and set the DIP switch (see page
50) before connecting the IX2-UCB control board to the IX81.

This microscope employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only
with modules designed for the IX2 or BX2 series (which belong to the Olympus IX or BX series).
For the applicable modules, please consult Olympus or the latest catalogues. Less than optimum per-
formance may result if inappropriate module combinations are used.

Configuration of Instruction Manuals

Since this microscope is expandable to a variety of systems, separate instruction manuals are prepared so
that the user has to read only the manuals according to the user’s own system.

Manual Name

Main Contents

IX81

Observation procedures including transmitted light brightfield,
phase contrast and DIC observations

IX2-UCB/U-HSTR2

Functions of the Control Box (incorporating the power supply)
and Hand Switch

IX2 Software for PC (CD-ROM)

*Model name IX2-BSW (Ver. 01.03 or later)

Methods of PC control of functions

Fluorescence system for IX2

Reflected light fluorescence observation

U-FWT/FWR/FWO

Motorized filter wheels (The U-FWT cannot be used with this microscope.)

Fig. 1

Precautions When Unpacking the Microscope

Releasing the Transport Lock of the Motorized Revolving Nosepiece

# Never attempt to turn the system on without removing the clamping

rod. Otherwise, the focusing mechanism may be damaged.

1. Loosen the screw @ of the clamping rod using the Allen screwdriver

provided with the microscope frame.

2. Turn the nut ² of the clamping rod in the direction of the arrow until the

nut reaches the lowest height.

3. Likewise, turn the clamping rod ³ in the direction of the arrow until the

rod reaches the lowest height.

4. Insert the Allen screwdriver into the hole on the clamping rod mount 4

and turn the screwdriver in the opposite direction to the dial to remove
the clamping rod.

5. Attach the provided seal (10 mm dia., white) on the hole made after

removing the clamping rod to prevent penetration of dust through the
hole.

}Retain the clamping rod carefully because it will be used again the next

time the microscope is transported.

|

@

3

²

5

IX2-MLWCD

IX2-DICD

DIC condenser

Mid long working distance condenser