OTHER OBSERVATION METHODS
5-1 Phase Contrast Observation
}A phase contrast objective, phase contrast optical element and the U-CT30 centering telescope are required for phase
contrast observation (the U-BI90CT binocular observation tube .is not required).
}If a DIC slider, analyzer or polarizer is engaged in the light path, disengage it.
1
Phase Contrast Optical Elements and Applicable Objectives
With the IX2-LWUCDA2 and IX2-LWUCD Condenser
Optical Element
Indication
Applicable Objectives
IX-PHL (small)
PhL
UPlanFl4XPh
IX-PHC (small)
PhC
CPlan10XPh, LCAch20XPh, CPlanFl10XPh
IX-PH1 (small)
Ph1
UPlanFl10XPh, UPlanFl20XPh, LCPlanFl20XPh, UPlanApo10XPh
IX-PH2 (small)
Ph2
UPlanFl40XPh, LCPlanFl40XPh, LCPlanFl60XPh, UPlanApo20XPh, LCAch40XPh,
LUCPlanFl40XPh, SLCPlanFl40XPh
IX-PH3 (large)
Ph3
UPlanFl100XO3Ph, UPlanApo40XO
I3Ph, UPlanApo100XOI3Ph, PlanApo60XO3Ph
}Insert the optical element (small) in the 30 mm position and the optical element (large) in the 38 mm position.
For well observation, it is recommended to use the IX-PHC to obtain the phase contrast effect in a wide range of field of
view.
Optical Element
Indication
Applicable Objectives
IX-PHL (small)
PhL
UPlanFLN4XPh
IX-PHC (small)
PhC
CPlanN10XPh, LCAchN20XPh, CPlanFLN10XPh
IX-PH1 (small)
Ph1
UPlanFLN10X2Ph, UPlanFLN20XPh, LUCPlanFLN20XPh
IX-PH2 (small)
Ph2
UPlanFLN40XPh, LUCPlanFLN40XPh, LUCPlanFLN60XPh,
LCAchN40XPh, UPlanFLN60XOIPh
IX-PH3 (large)
Ph3
UPlanFLN100XO2Ph
UIS2 Series
UIS Series