beautypg.com

3 simplified polarized light observation – Olympus IX81 User Manual

Page 42

background image

37

Fig. 52

5-3 Simplified Polarized Light Observation

}An analyzer and polarizer are required for simplified polarized light observation. The applicable condenser is the

IX2-LWUCDA2 or IX2-LWUCD. Simplified polarized light observation is not available with the IX-ULWCD.

1

Attaching the Analyzer and Polarizer

}Use the same procedure as that for attaching the analyzer and polarizer

for DIC observation (see page 34).

2

Observation Method

(Fig. 52)

1. Rotate the turret for the BF (brightfield) light path (with no optical element

engaged in the light path).

2. Move the polarizer detaching lever 1 on the IX-LWPO polarizer unit to

engage the polarizer in the light path. (Fig. 52)

3. Attach the objective to be used to the revolving nosepiece and rotate it to

engage the objective in the light path.

4. Move the polarizer rotation/clamping knob 2 on the polarizer unit hori-

zontally in the counterclockwise direction around the axis until the posi-
tion with which the field of view is darkest. (Fig. 52)

5. Clamp the polarizer rotation/clamping knob so that the polarizer will not

rotate.

6. Place a specimen on the stage and bring the specimen into focus. Now

the specimen can be observed with simplified polarized light observa-
tion.

7. Adjust the field iris diaphragm so that its image circumscribes the field of

view.

8. Adjust the aperture iris diaphragm to enhance the contrast.

5-4 Reflected Light Fluorescence Observation (Separate Manual)

}Refer to the separate instruction manual.

1

2