beautypg.com

Assembly, 1 assembly diagram – Olympus IX81 User Manual

Page 54

background image

49

ASSEMBLY

9-1 Assembly Diagram

The diagram below shows the sequence of assembly of the modules. The numbers indicate the order of assembly.
The module numbers shown in the following diagram are merely the typical examples. For the modules with which the
module numbers are not given, please consult your Olympus representative or the latest catalogues.

# When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts

or touching glass surfaces. Also be sure to release the transport lock of the revolving nosepiece and light path
selector by removing the screws before use (see pages 1 & 2).
Assembly steps enclosed in will be detailed on the subsequent pages.

}All assembly operations are possible by using the Allen screwdriver (with across flats of 3 mm, ) and Allen wrench

(width across flats of 4 mm, ) provided with the microscope. For assembly of the reflected fluorescence system
and modules without descriptions, refer to the separate instruction manuals.

Terasaki plate holder
IX-HOT
Slide glass holder
IX-HOS
Petri dish holder
IX-HOP

}For the connection of cables, refer to the instruction manual for the IX2-UCB control box.

Mechanical stage
IX-MVR

Stage Center
plate

Plain stage
IX2-SP

Cross stage
IX-SVL2

Cross stage
IX2-SFR

UCD adapter
IX-ADUCD

Universal
condenser
U-UCD8

Universal
condenser
IX2-LWUCD

Motorized
universal
condenser
IX2-LWUCDA2

Condenser
IX-ULWCD

Objective

Illumination
column
IX2-ILL100

Halogen lamp housing
U-LH100L-3
U-LH100-3
IX-HLSH100

Halogen bulb
12V100WHAL-L

High-intensity light
source
U-LH100HG
U-LH100HGAPO
U-LH75XEAPO
U-LH50MH

Critical
illumination
lens
IX-CI

Filter wheel
U-FWR
Motorized shutter
IX2-SHA

Eyepieces
WHN10X
WHN10X-H

Trinocular
observation tube H
U-TR30H-2

Binocular
observation tube
U-BI90CT
U-BI90
U-TBI90

Filter wheel
U-FWO

Intermediate tube
IX-ATU

Intermediate tube
U-EPA2
U-CA
IX-SDO, etc.
See page 4 to 5 for the restrictions.

Microscope frame
IX81S1F-3
IX81S8F-3
IX81F-3

Microscope frame
clamping plates
IX2-FP

Focus adjustment
knob interface
U-IFFH

Focus adjustment
knob unit
U-FH

Hand switch
U-HSTR2

Control box
IX2-UCB

PC

Z-Board
U-ZPCB(T2)

Medium
working
distance
condenser
IX2-MLWCD