PHOTOMICROGRAPHY AND TV OBSERVATION

6-1 Photomicrography

}Use the U-TR30H-2 trinocular observation tube or the side port for photomicrography.

Photomicrography can be performed using the PM-10, PM-20 or PM-30 photomicrographic system.
For how to use the photomicrographic system, see the instruction manual of the photomicrographic system in use. The
following are procedures related to this microscope.

1

When the U-TR30H-2 trinocular observation tube is used
(in combination with the IX-SPT)

(Fig. 53)

3

2

1

Attaching the Straight Photo Tube
(IX2-SPT/IX-SPT)

(Figs. 53 & 54)

Fig. 54

Fig. 55

When the side port is used (in combination with the IX2-SPT) (Fig. 54)

1. Using the Allen screwdriver, loosen the side port clamping screw 1 of

the microscope frame.

2. Align index 2 on the IX2-SPT straight photo tube with positioning index

3 on the side port and fit the straight photo tube.

3. Tighten the clamping screw 1 firmly.

2

Attaching the Photo Eyepiece

(Fig. 55)

Use only a PE photo eyepiece for photomicrography.
Insert the PE photo eyepiece 1 into the photographic system mount on
the straight photo tube.

}The IX2-SPT or IX-SPT is equipped with a plunger for use in preventing

movement of the PE photo eyepiece. Be sure to push in the plunger.

# The U-SPT straight photo tube cannot be mounted.
1. Using the Allen screwdriver, loosen the clamping screw 1 on the photo

port of the trinocular observation tube and remove the cap.

2. Align index 2 on the front pf the photo port of the trinocular observation

tube with index 3 on the IX-SPT straight photo tube, then fit the circular
dovetail of the straight photo tube into the photo port of the trinocular
observation tube.

3. Tighten the clamping screw 1 firmly.

1

2

3

1