Viewing/setting replenishment settings, Viewing the replenishment rate, Setting the chemical replenishment rate – Kodak PROFESSIONAL LED II 20P User Manual
Page 68: Setting the wash replenishment rate, Calibrating the replenishment flow rate, Measuring the flow rate, Viewing/setting replenishment settings -36, Calibrating the replenishment flow rate -36, Measuring the flow rate -36
System Operating Procedures
3-36
September 2000
Viewing/Setting Replenishment Settings
Viewing the Replenishment Rate
1.
Press Replenishment to access the Replenishment screen.
2. Press
the
Up or Down Arrow to select the desired replenishment pump or the
wash/stabilizer tanks.
Setting the Chemical Replenishment Rate
1.
Select the desired replenishment pump.
2. Press
Set to access the Set mode.
3. Use
the
Up or Down Arrow to change the replenishment needed for each
square foot of paper processed.
4. Press
Set twice or press a different blue function key to return to View mode.
NOTE: Press Replenishment to toggle the pump to Manual on, Manual off or
Per-Square-Foot mode.
Setting the Wash Replenishment Rate
1.
Select the desired wash/stabilizer tank.
2. Press
Set to access the Set mode.
3. If the wash is in per-square-foot mode, use the Up or Down Arrow to change
the amount of wash solution needed for each square foot of paper processed.
NOTE: In Automatic mode, the wash solution is on whenever material is in the
processor.
4. Press a different blue function key to return to View mode.
NOTE: Press Replenishment to toggle the pump to Manual on, Manual off,
Per-Square-Foot, or Automatic mode.
Calibrating the Replenishment Flow Rate
NOTE: Turn the key switch to the service position before you calibrate the
replenishment flow rate.
IMPORTANT: Before calibrating a flow rate, position the desired replenishment
tube into a beaker to catch the pump’s output or the wash tank’s
overflow. Make sure the beaker is large enough to catch the output
of the pump for at least 10 replenishment bursts of one second
duration. The pump may make more than one revolution during a
single burst period.
Measuring the Flow Rate
1.
Select the desired replenishment pump.
2. Press
Set twice to access the Replenishment screen in the Calibrate mode.
3. Press
Replenishment to initiate a 10-second auto calibrate cycle for the
selected pump. After the cycle has completed, measure the output and divide
it by ten. This is the actual flow rate.