Specifications

September 2000

B-3

Plumbing

NOTE: Always adhere to local ordinances and regulations for wastewater disposal.

A plumbed processor with a water wash system is required for environments with
temperatures above 27

o

C (80

o

F).

Plumbed Processors – Water Supply

You are responsible to provide adequate water supply and drainage for the
processor before it is installed.

We recommend that you order a water supply panel for plumbed systems from
Kreonite Inc. The water supply panel must be installed by Kreonite. However, a
Kodak representative will install flexible supply line to the processor and an
electrical connection to the solenoid. The water supply panel should be wall-
mounted within 4.9 m (16 ft.) of the rear of the processor.

Installation, technical support and warranty of the water supply panel is the
responsibility of Kreonite.

Water to a plumbed system must meet the following specifications:

Flow Rate:

3.8 – 20.0 liters per minute (1 – 5 gallons per minute)

Temperature: Hot – 59

o

C (120

o

F) minimum

Cold – 27

o

C (80

o

F) maximum

Plumbed Processors – Waste Water and Effluents

A 2 inch floor drain should be located near the processor. The drain system
should conform to the local specifications for silver-laden bleach-fix and stabilizer
or waste water.

A Kodak representative will install a flexible drain tube to the processor. You are
responsible for installing the flexible drain tube to a locally approved drain system.
For unplumbed systems,10-litre containers are installed under the processor to
transport waste to the site’s waste facilities.

We recommend that you install a 61 cm (24 in.) wide sink near the processor for
cleanup and processor maintenance.

Unplumbed Processors

We recommend that you install a 61 cm (24 in.) wide sink near the processor for
cleanup and processor maintenance.