Head – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 95

95
Head
ThinSliceSeq
RoutineSeq
kV
120
120
Effective mAs
250
250
Slice collimation 0.75 mm
1.5 mm
Slice width
4.5 mm
9.0 mm
Feed/Scan
9.0 mm
18.0 mm
Rotation time
0.5 sec.
0.5 sec.
Kernel
H31f
H31f
Image order
ca-cr
ca-cr
CTDI
w
58.2 mGy
60.3 mGy
Effective dose
Male: 0.7 mSv
Male: 1.6 mSv
Female: 0.8 mSvFemale: 1.8 mSv
PosteriorFossaOptimization Filter
PFO: To reduce beam-hardening artifacts in head
images, particularly in the base of the skull, use the
Posterior Fossa Optimization (PFO) filter.
If you want to set up the scan with only one range,
i. e. either 12 x 0.75 mm or 12 x 1.5 mm, you can
easily delete one mode by clicking the chronicle with
the right mouse button, and select cut.
Contrast medium IV injection
Start delay
60 sec.
Flow rate
2 ml/sec.
Total amount
50 – 60 ml