Headspi0.5s, Children – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 72
72
HeadSpi0.5s
Indications:
Spiral mode for routine head studies, e. g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.
HeadSpi0.5s
kV
120
Effective mAs
*
Slice collimation
1.5 mm
Slice width
4.0 mm
Feed/Rotation
12.2 mm
Rotation time
0.5 sec.
Kernel
C30f
Increment
4.0 mm
Image order
ca-cr
Contrast medium IV injection
Start delay
exam dependent
Flow rate
dependent upon needle size/Access site
Total amount
1– 2 ml per kg of body weight
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old, and a scan
range of 90 mm was used.
** The conversion factor for a 7-year-old, 260 mAs and
a scan range of 110 mm was used.
Body Weight kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 6 months
120 90
17.1
Male: 1.7*
Female: 1.7*
6 months –
120 150
28.5
3 years
3 years –
120 220
41.8
Male: 4.4**
6 years
Female: 4.3**
Children