beautypg.com

Headspi0.5s, Children – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 72

background image

72

HeadSpi0.5s

Indications:

Spiral mode for routine head studies, e. g. tumors,
hydrocephalus, hemorrhaging, abnormalities, etc.

HeadSpi0.5s

kV

120

Effective mAs

*

Slice collimation

1.5 mm

Slice width

4.0 mm

Feed/Rotation

12.2 mm

Rotation time

0.5 sec.

Kernel

C30f

Increment

4.0 mm

Image order

ca-cr

Contrast medium IV injection
Start delay

exam dependent

Flow rate

dependent upon needle size/Access site

Total amount

1– 2 ml per kg of body weight

* The mAs should be adjusted to the age of the child.

* The conversion factor for an 8-week-old, and a scan

range of 90 mm was used.

** The conversion factor for a 7-year-old, 260 mAs and

a scan range of 110 mm was used.

Body Weight kV

mAs

CTDI

w

Effective dose

(mGy)

(mSv)

< 6 months

120 90

17.1

Male: 1.7*
Female: 1.7*

6 months –

120 150

28.5

3 years
3 years –

120 220

41.8

Male: 4.4**

6 years

Female: 4.3**

Children