Children carotidcta – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 78
78
Children
CarotidCTA
Indications:
CT angiography of the carotid arteries, e. g. carotid
stenosis or occlusion, vascular abnormalities of the
carotids or vertebral arteries, etc.
CarotidCTA
2
nd
Reconstr.
kV
80
Effective mAs
*
Slice collimation
0.75 mm
Slice width
5.0 mm
1.0 mm
Feed/Rotation
12 mm
Rotation time
0.5 sec.
Kernel
B30f
B20f
Increment
5.0 mm
0.7 mm
Image order
cr-ca
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old, and a scan
range of 80 mm was used.
** The conversion factor for a 7-year-old and a scan
range of 120 mm was used.
Age
kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 6 years
80
100
2.6
Male: 0.6*
Female: 0.7*
6 years –
80
150
3.9
Male: 0.4**
12 years
Female: 0.4**