beautypg.com

Abdomen – Siemens SOMATOM SENSATION CARDIAC A60 User Manual

Page 121

background image

121

AngioRoutine

2

nd

Reconstr.

kV

120

Effective mAs

130

Slice collimation

0.75 mm

Slice width

5.0 mm

1.0 mm

Feed/Rotation

15.0 mm

Rotation time

0.5 sec.

Kernel

B30f

B20f

Increment

5.0 mm

0.7 mm

Image order

cr-ca

CTDI

w

10.1 mGy

Effective dose

Male: 5.8 mSv
Female: 7.5 mSv

Do not administer oral contrast medium, as this
impairs the editing of MIP/SSD/VRT images.

* CARE Bolus may be used to optimize the bolus timing.

Set the ROI for monitoring scan in the abdominal
aorta at the beginning of the scan range with trigger-
ing threshold of 120 HU, or use manual triggering.

MIPthin: AngioRoutine

sagittal

Image thickness

3

Distance between images

3

Number of images

60

For the 2

nd

reconstruction the Autoload into the

MIPthin Range on the 3D Card is activated. The images
will be automatically loaded into 3D, MIPthin, and
sagittal MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.

Contrast medium IV injection
Start delay

20 – 25 sec.*

Flow rate

3.0 – 3.5 ml/sec.

Total amount

100 – 120 ml

Abdomen