Thorax – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 111

111
Thorax
ThoraxRoutine
kV
120
Effective mAs
100
Slice collimation
1.5 mm
Slice width
6.0 mm
Feed/Rotation
30.0 mm
Rotation time
0.5 sec.
Kernel
B41f
Increment
6.0 mm
Image order
cr-ca
CTDI
w
7.0 mGy
Effective dose
Male: 2.8 mSv
Female: 3.6 mSv
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
Contrast medium IV injection
Start delay
25 – 30 sec.
Flow rate
2.5 ml/sec.
Total amount
80 ml