HeartView CT
CaScoreSeqStd
Indications:
This is a sequential scanning protocol using an ECG
triggering technique for coronary calcium scoring
studies.
Topogram:
AP, 512 mm.
From the carina until
the apex of the heart.
CaScoreSeqStd
kV
120
Effective mAs
30
Slice collimation
1.5 mm
Slice width
3 mm
Feed/Rotation
18 mm
Rotation time
0.42 sec.
Temporal resolution
210 ms
Kernel
B35f
Image order
cr-ca
CTDI
w
2.2 mGy
Effective Dose
Male: 0.5 mSv
Female: 0.7 mSv
If you apply API for image acquisition, please make
sure that the breathhold interval in the Patient Model
Dialog is longer than the total scan time, e. g. 50 s,
otherwise the image acquisition will be interrupted by
the default breathhold interval. This does not apply