Heartview ct coronarystd – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 37
37
HeartView CT
CoronaryStd
Indications:
This is a standard spiral scanning protocol, using a
Rotation Time of 0.42 s, with an ECG gating technique
for coronary CTA studies.
Topogram:
AP, 512 mm.
Approximately, from
the carina until the
apex of the heart.
A typical range of
12 cm covering the
entire heart can
be done in 18.7 s.
CoronaryStd
kV
120
Effective mAs
500
Slice collimation
0.75 mm
Slice width
1 mm
Feed/Rotation
2.8 mm
Rotation time
0.42 sec.
Temporal resolution
Up to 105 ms*
Kernel
B30f
Increment
0.5 mm
Image order
cr-ca
CTDI
w
42.0 mGy
Effective dose
Male: 7.0 mSv
Female: 10.2 mSv
* Depends on heart rate.