Children – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 77

77
Tips
• Children, who are more than 12 years old should be
scanned with an adult protocol.
• The first recon job is defined for soft tissue visuali-
zation.
Contrast medium IV injection
Start delay
18 – 20 sec.
Flow rate
dependent upon needle size/Access site
Total amount
1– 2 ml per kg of body weight
MIPthin: HeadAngio
coronal
Image thickness
3
Distance between images
3
Number of images
35
For the 2
nd
reconstruction the Autoload into MIPthin
Range on the 3D Card is activated. The images will be
automatically loaded into 3D, MIPthin, and a coronal
MIPthin Range will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Children