HeartView CT
CorStd_LowHeartRate
Indications:
This is a special spiral scanning protocol for coronary
CTA studies. It uses ECG gating technique and a 0.42 s
rotation time, and should be used for patients with
heart rate below 50 bpm.
Topogram:
AP, 512 mm.
Approximately, from
the carina until the
apex of the heart.
A typical range of
12 cm covering the
entire heart can
be done in 23.7 s.
CoronaryStdLow
kV
120
Effective mAs
500
Slice collimation
0.75 mm
Slice width
1.0 mm
Feed/Rotation
2.2 mm
Rotation time
0.42 sec.
Temporal resolution
Up to 105 ms*
Kernel
B30f
Increment
0.5 mm
Image order
cr-ca
CTDI
w
42.0 mGy
Effective dose
Male: 7.0 mSv
Female: 10.2 mSv