Heartview ct coronarycare0.5s – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 41
41
HeartView CT
CoronaryCARE0.5s
Indications:
This is a spiral scanning protocol, using a Rotation
Time of 0.5 s, ECG gating technique and a dedicated
cardiac filter which can reduce images noise thus
makes the dose reduction possible for coronary CTA
studies.
Topogram:
AP, 512 mm.
Approximately, from
the carina until the
apex of the heart.
A typical range of
12 cm covering the
entire heart can
be done in 18.4 s.
CoronaryCARE0.5s
kV
120
Effective mAs
267
Slice collimation
0.75 mm
Slice width
1 mm
Feed/Rotation
3.4 mm
Rotation time
0.5 sec.
Temporal resolution
Up to 125 ms*
Kernel
B30f
Increment
0.5 mm
Image order
cr-ca
CTDI
w
22.4 mGy
Effective dose
Male: 3.8 mSv
Female: 5.5 mSv
* Depends on heart rate.