Headangio, Children – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 76

76
HeadAngio
Indications:
Spiral mode for head CT angiography, e. g. cerebral
vascular abnormalities, tumors etc.
HeadAngio
2
nd
Reconstr.
kV
80
Effective mAs
*
Slice collimation
0.75 mm
Slice width
2.0 mm
1.0 mm
Feed/Rotation
6.8 mm
Rotation time
0.5 sec.
Kernel
H20f
H10f
Increment
2.0 mm
0.7 mm
Image order
ca-cr
* The mAs should be adjusted to the age of the child.
* The conversion factor for an 8-week-old, and a scan
range of 40 mm was used.
** The conversion factor for a 7-year-old and a scan
range of 60 mm was used.
Age
kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 6 years
80
100
8.4
Male: 0.4*
Female: 0.4*
6 years –
80
150
12.6
Male: 0.3**
12 years
Female: 0.4**
Children