Heartview ct – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 27
27
HeartView CT
Absolute – delay: a fixed time delay after the onset of
the R-wave (Fig. 11).
Absolute – reverse: a fixed time delay prior to the onset
of the next R-wave (Fig. 12).
Preview Series Reconstruction
Preview series can be used to define the optimal time
window before the full series is reconstructed.
Click on the preview series button in the Trigger card.
The slice position of the preview series is based on the
currently displayed image in the tomogram segment,
which has to be chosen by the user.
Fig. 11
Time
Estimated
R-Peak
ECG
(t)
- 400 msec
Scan/
Recon
Fig. 12
Time
ECG
(t)
400 msec
Scan/
Recon