Children abdcta – Siemens SOMATOM SENSATION CARDIAC A60 User Manual
Page 86

86
Children
AbdCTA
Indications:
For abdominal CT Angio studies, e. g. vascular
abnormalities, aneurysms, etc.
AbdCTA
2
nd
Reconstr.
kV
80
Effective mAs
*
Slice collimation
0.75 mm
Slice width
3.0 mm
1.0 mm
Feed/Rotation
12 mm
Rotation time
0.5 sec.
Kernel
B30f
B20f
Increment
3.0 mm
0.7 mm
Image order
cr-ca
* The mAs should be adjusted to the body weight
of the child.
* The conversion factor for an 8-week-old, and a scan
range of 100 mm was used.
** The conversion factor for a 7-year-old and a scan
range of 200 mm was used.
Body weight
kV
mAs
CTDI
w
Effective dose
(mGy)
(mSv)
< 15 kg
80
30
0.8
Male: 0.4*
Female: 0.6*
15 – 24 kg
80
50
1.3
25 – 34 kg
80
75
1.9
Male: 0.9**
Female: 1.4**
35 – 44 kg
80
110
2.9
45 – 54 kg
80
150
3.9