Industrial Data Systems IDS DT220 User Manual
Page 21

DT220 Users Manual version 1.a Configuration Parameters
17
Scale Parameter 6.3: Zero tracking band (12)
The zero tracking band is the maximum displacement from zero wherein zero tracking is
active. In order for zero tracking to be active, the following conditions must be met:
• Scale weight is within Zero Track Band.
• Scale is stable (no motion)
• Zero Track Band is greater than 0 (zero).
The zero tracking bands are entered in units of ¼ dd, i.e.. For a band of:
½ display division enter 2
½ x 4 = 2.
2 display divisions enter 8
2 x 4 = 8.
4.5 display divisions enter 18
4.5 x 4 = 18
Scale Parameter 7.1: Motion Detection Delay (6)
Motion detection delay is used by the motion detection function to determine if the scale
is in motion (not stable). Enter the amount of time (in tenths of a second) that the scale
must be stable for motion status to be false (scale status = stable).
Example:
0 = disables motion detection.
2 = 0.2 second delay before motion = false (stable = true).
15 = 1.5 second delay.
30 = 3.0 second delay.
Scale Parameter 7.2: Motion Detection Band (6)
The motion band is used by the motion detection function to determine if the scale is in
motion (not stable). If the amount of change between display updates (parameter 2) is
greater than the motion band, a MOTION status is generated (unstable condition). The
motion band is entered in units of ¼ dd, i.e.. For a motion band of:
½ display division enter 2
½ x 4 = 2.
2 display divisions enter 8
2 x 4 = 8.
4.5 display divisions enter 18
4.5 x 4 = 18.
Scale Parameter 7.3: Display Blanking During Motion (No)
When display blanking is enabled, the weight display is blanked (turned off) when the
scale is in motion (not stable). Enter a 1 (yes) to enable display blanking, enter a 0 (no) to
disable display blanking.