beautypg.com

Source output accuracy 6-2, Setup: 6-2, Test procedure 6-2 – Anritsu ME7840A User Manual

Page 98: Test equipment 6-1, Chapter 6 performance verification procedure, Introduction -1, Conventions -1, Test equipment -1, Source output accuracy performance verification, Protection replace

background image

6-4

SOURCE OUTPUT
ACCURACY

This test verifies the source output accuracy of the MS462XC. The test
procedure uses the Power ALC Verification built-in function of the
MS462XC Basic Measurement software.

Setup:

Step 1.

Connect a GPIB cable between the power meter GPIB connec-
tor and the MS462XC dedicated GPIB connector, as shown in
Figure 6-1.

Step 2.

Turn on the MS462XC VNMS and allow it to warm-up at
least one hour.

Test Procedure

Step 1.

Disconnect cables between MS462XC

Port 1

and

Port

1 of

MS4782 Test Set and between

Port 3

of MS462XC and

Port 3

of MS4782 Test Set.

Step 2.

Press

Utility

key and select:

DIAGNOSTICS

TROUBLESHOOTING

MORE

VERIFY ALC CALIBRATON

Step 3.

Calibrate and zero the power sensor.

SOURCE OUTPUT ACCURACY

PERFORMANCE VERIFICATION

6-2

ME 7840A OMM

Noise

In

Noise

+28V

NO OPERATOR SERVICE-
ABLE PARTS INSIDE.
REFER SERVICING TO
QUALIFIED PERSONNEL.

WARNING

CAUTION

FOR CONTINUED FIRE

PROTECTION REPLACE

ONLY WITH SPECIFIED

TYPE AND RATED FUSE.

IEEE 488.2

GPIB

Dedicated

GPIB

Serial Port

Ethernet

VGA

SCSI-2

Ext

Analog
Output

Ext

Trigger

LPT Line Printer

Ext I/O

Port 1

Port 2

Bias

Bias

Port 2

Bias Fuse

Port 1

Bias Fuse

Ext Analog

In

Ext

Source

Ext 10 MHz

In

47-440 Hz

85-264VAC

REPLACE FUSE ONLY WITH

SAME TYPE AND RATING

CAUTION

DO NOT OPERATE

WITH POWER CORD

UNGROUNDED

N274

00000001

Made in UK

12-24V DC
(1.3A)

OUTPUT 1

OUTPUT 2

INPUT 1
DIGITAL

INPUT 2
ANALOG

CALIBRATOR

B

RS 232

PARALLEL PRINTER

GPIB / IEEE 488

CAUTION

DO NOT OPERATE

WITH UNGROUNDED

POWER CORD.

85-250 VAC
47-440Hz

40VA MAX

SEE MANUAL FOR VOLTAGE RANGES ON BNC CONNECTORS

A

NO OPERATOR SERVICE-

ABLE PARTS INSIDE.

WARNING

Figure 6-1. ML2430A Series Power Meter GPIB connection to the MS462XX Vector Network Measurement System