beautypg.com

Anritsu ME7840A User Manual

Page 106

background image

Step 25.

Press

Power

key and select

SOURCE 1 SETUP

.

Step 26.

Set

PORT 1 ATTN

to

10 dB.

Record the Power Meter Readout

and the Marker 1 Readout to the table on next page. Repeat
this step for other

PORT 1 ATTN

setting listed in the table.

Step 27.

Verify if the difference of the two measured values are within
specifications (Table 6-2).

RECEIVER DISPLAY LINEARITY

PERFORMANCE VERIFICATION

6-10

ME 7840A OMM

PORT 1 ATTN

Setting (dB)

Power Meter

Reading

MS462XC Marker 1 Reading

Difference Between Two

Readings

Specification

10

£

0.1 dB

20

£

0.1 dB

30

£

0.1 dB

40

£

0.2 dB

50

£

0.2 dB

60

£

0.2 dB

Table 6-2. Receiver Magnitude Display Linearity Specifications