beautypg.com

Cal files setup.txt a-11 - 12, Power level calibration specification file a-11, Power level.txt a-11 – Anritsu ME7840A User Manual

Page 122: A-12)

background image

CAL FILES SETUP.TXT

CAL SPECIFICATION FILES

A-12

ME 7840A OMM

ndut,1,

//Number of DUTs

type,powsweep

type,harm

type,imd

POWSWEEP_START

ncal,1,

//Number of Power Sweep calibrations

npts,3,

//Number of points (default for all)

if,3000,

//IF default

avgs,20,

//averages default

pow,-15,

//powerlevel start default

switch,8,

//switch setting

CAL,1, name,PowSweep,//File name root (append _1 for DUT1, _2 for

DUT2, etc

f1,900,

//Main Tone MHz

f2,931,

//Offset KHz

POWSWEEP_END

HARM_START

ncal,1,

//Number of Harmonic calibrations

if,3000,

//IF default

avgs,20,

//averages default

pow,0,

//power level start default

switch,8,

//switch setting

CAL,1, name,Harmonics,//File name root (append _1 for DUT1, _2 for DUT2,

etc

f1,900

//Main Tone MHz

f2,3000,

//Max Range MHz

HARM_END

IMD_START

ncal,1,

//Number of IMD calibrations

npts,4,

//

pow,-15,

//power level setting dBm

if,30000,

//IF Bandwidth

avgs,250,

//Averages

switch,8,

//switch setting

CAL,1, name,Imd,

//File name root (append _1 for DUT1, _2 for DUT2,

etc

f1,900,

//Main Tone MHz

f2,977,

//Offset KHz

IMD_END

Figure A-2. Cal Files Setup Calibration Specification File

{

Header

Calibration

Specifications

{

Section

1

{

Section

2

{

Section