Imd 5-26, Imd operations, measurement – Anritsu ME7840A User Manual

Page 87

background image

5-9

IMD

This test measures inter-modulation distortion.

Procedure

Prepare the ME7840A as shown in Chapter 3.

Enter the appropriate test values in the displayed
User Input screen. The input fields are described
below:

n

Measurement Type: Already set to “IM” for IMD

– input disabled.

n

Main Tone (MHz): CW Frequency setting for

source 1.

n

Offset (kHz): Offset frequency for Source 2

(Source 2 will be set to Main Tone + Offset).

n

Input Power (dBm): Input RF power to AUT.

n

IMD Order:NEED

n

Test Set Setting: Byte value for setting test set

attenuation and source switch:

IMD

OPERATIONS, MEASUREMENT

5-26

ME 7840A OMM

Internal Source

Value

External Source

Value

0 dB

8

0 dB

0

10 dB

9

10 dB

1

20 dB

10

20 dB

2

30 dB

11

30 dB

3

40 dB

12

40 dB

4

50 dB

13

50 dB

5

60 dB

14

60 dB

6

70 dB

15

70 dB

7