sure that the equipment is properly connected to the bus. If
so, the following screen will be displayed:

Step 3.

Click on “Run Calibration Procedure” to display a list of cali-
bration specification files (below). (Refer to paragraph 4-3 for
a description of these files.) Running the files is described in
succeeding numbered paragraphs.

4-6

POWER LEVEL

The A_Power Level.txt specification file provides prompts for a power
level calibration. Proceed as follows:

Step 1.

Zero-set the power sensor for accurate measurement in fol-
lowing steps.

Step 2.

Select “Tools” then “Run Calibration Procedure.”

POWER LEVEL

OPERATIONS, CALIBRATION

4-4

ME 7840A OMM