Technical details cal specification files – Anritsu ME7840A User Manual
Page 118
TECHNICAL DETAILS
CAL SPECIFICATION FILES
A-8
ME 7840A OMM
3.
powsweep (Power Sweep
-
IMD mode or 2-tone mode):
Header Section Keywords:
ncal
Number of calibration sequences of this type to run.
npts
NOTE: npts is IGNORED. Always uses 3, which refers to number of frequency values
allowed (must be 3 in current software).
if
IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000
NOTE: Illegal value causes software to default to 1000.
avgs
Averages: (Optional). Software will set averaging to value (and turn on) during cal.
pow
Start power level for sweep: Set on for Source 1 and Source 2.
switch
Setting for test set. Correponds to 0
-
15 settings.
Detail Section Keywords:
CAL, 1
First calibration
name
Name of *.cal file to save when calibration sequence completes.
f1
Start frequency in MHz for global calibration
NOTE: Set a wide frequency range. Miminum value is 10; ensure f1 is less
than 2.
f2
Stop frequency in MHz.
Example
Refer to Figure A-2 (page A-12).
Table A-3. Technical Details of Calibration Specification Files (7 of 9).
Header
POWSWEEP_START
POWSWEEP_END
Details