Hot s22 4-8, Hot s22 operations, calibration – Anritsu ME7840A User Manual
Page 50

4-7
HOT S22
The “B_HotS22.txt” file provides prompts for setting up a Hot S
22
specifi-
cation file. Hot S
22
is a return loss measurement of an amplifier’s output
port (Port 2) while stimulus is applied to its input port. The prompts
cause the Scorpion to be placed into a state where it is ready to make cali-
brated measurements on the user’s AUT. At the completion of the pro-
gram, the state information is saved to a *.cal file. To run this program,
proceed as follows.
Step 1.
Select “Tools” then “Run Calibration Procedure.”
Step 2.
Double-click on “C_Hot S22.txt” to begin the Hot S
22
cal file
setup. Follow the prompt and connect open, short, and load to
Port 2 as directed.
HOT S22
OPERATIONS, CALIBRATION
4-8
ME 7840A OMM