beautypg.com

Receiver display linearity 6-8, Setup: 6-8 – Anritsu ME7840A User Manual

Page 104

background image

6-7

RECEIVER DISPLAY
LINEARITY

This test verifies the magnitude display linearity of the receiver.

Setup:

Step 1.

Setup the equipment as shown above.

Step 2.

Allow the system to warm up for at least one hour.

Step 3.

Press

Default

key then

0

key.

Step 4.

Press

Seq

key and select:

TTL I/O.

PARALLEL OUT SETUP

SET PARALLEL OUT PORT (0-255)

Step 5.

Change the value from 0 to 8.

Step 6.

Select:

RETURN

RETURN

Step 7.

Press

Freq

key.

Step 8.

Change

START FREQUENCY

to

800 MHz

and

STOP

FREQUENCY

to

1 GHz

.

RECEIVER DISPLAY LINEARITY

PERFORMANCE VERIFICATION

6-8

ME 7840A OMM

MS4782A

Test Set

ML243XA

Power Meter

MS462X3C

DRA Scorpion

Test

Port 1

Port 1

Port 1

Port 2

Port 2

Port 3

Port 3

Test

Port 2

50 Ohm Terminations

11N50B

POWER DIVIDER

POWER SENSOR

GPIB

34NN50A

Cal / Zero

ML2408A

Power Meter

Power Meter

ON / OFF

Sensor

Channel

Trigger

System

CALIBRATOR

0

1

4

7

2

5

8

+ / -

3

6

9

CLR

A

B

Figure 6-2. Receiver Magnitude Display Linearity Test Setup

NOTE

Zero and calibrate Power
Sensor prior to connecting
to the power divider.