TECHNICAL DETAILS
CAL SPECIFICATION FILES
A-6
ME 7840A OMM
2.
sp and Hot S
22
(S Parameters & Hot S
22
)
Header Section Keywords:
ncal
Number of calibration sequences of this type to run.
npts
Number of frequency points to use during the calibration sequence. Increasing
this number will increase calibration time.
Legal values are (3, 15, 51, 101, 201, 401, 801, 1601 only)
NOTE: Illegal value causes software to default to 51.
con
Connector sypes for Port 1 and Port 2: Options:
MN, FN - Male/Female N type
MK, FK - Male/Female K
MS, FS - Male/Female SMA
if
IF bandwidth: Legal values are 10, 30, 100, 300, 1000, 3000, 10000, 30000
NOTE: Illegal value causes software to default to 1000.
avgs
Averages: (Optional). Software will set averaging to value (and turn on) during cal.
pow
Source power setting: during cal, Source 1.
switch
Setting for test set. Correponds to 0
-
15 settings.
Detail Section Keywords:
CAL, 1
First calibration
cal type
Type of calibration to perform:
12 - 12 term
RR - Reflection Reverse
RF - Reflection Forward
8T - 1 Path 2 Port Forward
8R - 1 Path 2 Port Reverse
FT - Transmission Frequency Response
Name for Hot S
22
For Hot S
22
only. Value = offset in kHz of 2nd Source
Table A-1. Technical Details of Calibration Specification Files (5 of 9).
Header
SP_START
SP_END
Details