Chapter 2 Front Panel Operations

RIGOL

DSG3000 User’s Guide

2-43

Bus Trigger

After selecting “Bus” trigger, if the sweep mode is set to “Cont”, the

instrument starts a sweep each time the “*TRG” command is sent. If the

sweep mode is “Single”, press Single to meet the single sweep condition.

After that the instrument starts a sweep and then stops each time the

“*TRG” command is sent.

External Trigger

In external trigger, the RF signal generator accepts the trigger signal input

from the [TRRIGER IN] connector (as shown in the figure below) at the

rear panel. If the sweep mode is set to “Cont”, the instrument starts a

sweep each time a TTL pulse signal with specified polarity is received. If the

sweep mode is “Single”, press Single to meet the single sweep condition.

After that the instrument starts a sweep and then stops each time a TTL

pulse signal with specified polarity is received.

To specify the polarity of the TTL pulse signal, press Trig Slope to select

“Pos” or “Neg” and the default is “Pos”.