Level, Level -4 – RIGOL DSG3000 Series User Manual
Page 132

RIGOL
Chapter 6 Specifications
6-4
DSG3000 User’s Guide
Level
Setting Range
Specification level range Setting range
Maximum output
level
9kHz ≤ f <100kHz +7dBm
+10dBm
100kHz ≤ f <1MHz +13dBm
+15dBm
1MHz ≤f ≤3GHz
+13dBm
+25dBm
3GHz +13dBm +20dBm Minimum output level 9kHz ≤ f <100kHz -110dBm -120dBm 100kHz ≤ f ≤ 6GHz -130dBm -140dBm Setting resolution 0.01dB Absolute Level Uncertainty [1] Level uncertainty +13 to -60dBm -60 to -110dBm -110 to -130dBm 9kHz ≤f<100kHz ≤ 0.5dB (typ.) ≤ 0.7dB (typ.) 100kHz ≤f≤3GHz ≤ 0.7dB, ≤ 0.5 (typ.) ≤ 0.9dB, ≤ 0.5 (typ.) ≤ 0.7dB (typ.) 3GHz < f ≤ 6GHz ≤ 0.9dB, ≤ 0.5 (typ.) ≤ 1.1dB, ≤ 0.5 (typ.) ≤ 0.9dB (typ.) VSWR [2] 1MHz ≤ f ≤ 6GHz < 1.8 (typ.) Level Setting Setting time ALC state on, frequency fixed, temperature range: 20℃ to ≤ 5ms (typ.) Uninterrupted level setting range ATT fixed mode, ALC state on, level range -110dBm to +13dBm >20dB (typ.) Max. Reverse Power 50V 1MHz < f ≤ 6GHz 10W NOTE: [1] ALC state: on or auto mode, 20℃ to 30℃ [2] In 50Ω system, typical, level ≤-10dBm, ATT auto mode
30℃
Max. reverse power Max. DC voltage