beautypg.com

Basic operations, To output rf signal – RIGOL DSG3000 Series User Manual

Page 71

background image

RIGOL

DSG3000 Quick Guide

21

Basic Operations

To Output RF Signal

Output a RF signal with 1 GHz frequency and -40 dBm amplitude from the [RF

OUTPUT 50Ω] connector.

1. Restore to factory setting

Press System Reset Preset Type “Factory” and then press Preset to

restore the factory setting (by default, the frequency offset is 0 Hz and the

amplitude offset is 0 dB).

2. Frequency setting

Press FREQ Frequency, use the numeric keyboard to input the frequency
value (1) and select the desired unit (GHz) from the pop-up unit menu or unit

keys.

 The frequency units available are GHz, MHz, kHz and Hz.
 Pressing Enter can select the default unit Hz.
 You can also press the left/right direction key to enter the parameter editing

state and move the cursor to the specified digit; and then use the up/down

direction key or knob to modify the value.

 After the frequency is set, you can use the up/down direction key or knob to

modify the frequency at the current step.

 Press FREQ Frequency and then press Step to set the step.

3. Amplitude setting

Press LEVELLevel, use the numeric keyboard to input the amplitude value
(-40) and select the desired unit (dBm) from the pop-up unit menu or unit keys.
You can also press Enter to select the default unit dBm.

 The amplitude units available are dBm, -dBm, mV, uV and nV.
 You can also press the left/right direction key to enter the parameter editing

state and move the cursor to the specified digit; and then use the up/down

direction key or knob to modify the value.

 You can also use the up/down direction key or knob to modify the amplitude

at the current step.

 Press LEVELLevel and then press Step to set the step.

4. Enable the RF output

Press RF, the backlight goes on and the RF label is displayed in the status bar of