background image

RIGOL

10

DSG3000 Quick Guide

4. GPIB

It is used to connect the PC to realize remote

control.

5. OCXO (option)

OCXO is oven controlled crystal oscillator and

can provide frequency reference source with

higher temperature stability.

Note: Forty minutes of warm-up is required

for the OCXO to reach its rated frequency.

About the ordering information of the option,

please refer to the data sheet.

6. Power Input Connector, Fuse and Switch

Power input connector.

The AC power supplies supported by this

signal generator are 100-240 V, 45-440 Hz.

The maximum input power of the instrument

cannot exceed 60 W. When the signal

generator is connected to AC power supply

via this connector, the instrument selects the

correct voltage range automatically and users

do not need to select the voltage range

manually.

Fuse.

The fuse specification of this signal generator

is AC 250 V, T2A.

If the fuse needs to be replaced, please refer

to To Replace the Fuse.

Power Switch

It is used to turn on or off the signal

generator.