11

7. External Trigger Input Connector

When the trigger mode of Sweep or “Swp-Sine” of LF
output is “Ext”, this connector is used to input the

external trigger signal.

8. Signal Valid Output Connector

When the RF output frequency or amplitude is

modified, after certain response and processing time,

the internal circuit of the instrument outputs RF signal

with the specified frequency and amplitude via the RF

output connector at the front panel. During this

process, the [SIGNAL VALID] connector outputs a

pulse sync signal, indicating that the RF output signal

is valid.

High Level (3.3 V): indicate that the RF signal is in

configuration;

Low Level (0 V): indicate that the RF signal is

stable (namely, the signal is valid).

3.3V

0V

Power-on Time

RF Valid

Configuration

Time

Configuration

Time

RF Valid

… …

V

T

9. Sweep Output Connector

When the RF sweep function is enabled, the [SWEEP

OUT] connector outputs a signal (0 V to +10 V) while

the RF output connector at the front panel outputs the

sweep signal according to the current setting. Each

sweep corresponds to an output cycle.

When the sweep type is step, this output signal is

related to the sweep shape (“Triangle” or

“Ramp”) currently selected.

When the sweep type is list, the default of this

output signal is ramp.