To replace the fuse – RIGOL DSG3000 Series User Manual
Page 65
![background image](https://www.manualsdir.com/files/859156/content/doc065.png)
RIGOL
DSG3000 Quick Guide
15
To Replace the Fuse
To replace the fuse, please use the specified fuse and follow the steps below.
1. Turn off the instrument, cut off the power supply and remove the power cord;
2. Use a small straight screwdriver to prize out the fuse seat;
3. Take out the fuse;
4. Replace the old fuse with a specified fuse;
5. Re-install the fuse seat.
Figure 4 To Replace the Fuse
WARNING
To avoid electric shock, make sure that the instrument is turned off, the
power supply is disconnected and the fuse used is up to standard before
replacing the fuse.
The location of the
fuse seat
Fuse
Fuse Seat