beautypg.com

Fig.9, Maintenance 6. technical specifications – Sealey TA200 User Manual

Page 5

background image

4.8. Mass Air Flow (MAF) Sensor (fig 9)

4.8.1. A mass air flow sensor converts air flow readings into one of the following, depending on the

sensor:

a) a DC voltage

b) a low frequency signal

c) a high frequency signal

The TA200 is suitable for testing those sensors with a DC voltage or low frequency signal

output.

4.8.2. Connect the DC voltage type MAF sensor in the DC voltage testing fashion (see section 4.1)

and set the rotary switch on the meter to 20V DC. For a low frequency signal type MAF

sensor, connect the meter in the RMPx10 testing fashion (see section 4.7) and choose the

appropriate setting for number of cylinders.

4.8.3. Connect the black test lead probe to the ground terminal of the sensor and connect the red

one as illustrated in fig 9.

4.8.4. Switch on the meter and the vehicles ignition but

DO NOT start the engine.

4.8.5.

DC Voltage type sensor: The displayed value should usually be less than or equal to 1V

(consult the vehicle manufacturer’s specifications).

4.8.6.

Frequency type sensor: Refer to the vehicle manufacturer’s specifications to compare

against your readings.

4.9. Other functions

4.9.1. The TA200 can also be used to test the following automotive components:

Oxygen sensors, fuel injectors, temperature sensors, position sensors, MAP and baro

sensors.

4.9.2. Please refer to the vehicle manufacturer’s manual for specific test procedures.

WARNING! DO NOT attempt to repair or service your meter unless you are qualified to do so and have the relevant calibration,

performance test and service information.

5.1. General maintenance:

5.1.1. Periodically wipe the case with a damp cloth and mild detergent.

DO NOT use abrasives or solvents.

5.1.2. Clean the terminals with a cotton bud and mild detergent to prevent the build up of dirt/moisture in the terminals which may affect readings.

5.1.3. Turn the meter off when not in use and remove the battery if the meter is not to be used for some time.

5.2. To replace the fuses:

5.2.1. Turn the meter off and disconnect any leads/equipment from the terminals.

5.2.2. Remove the protective rubber jacket from the meter.

5.2.3. Remove the 3 larger screws from the rear of the case and open the meter.

5.2.4. Remove the fuse in question and replace with a fuse of identical type and specification:

Fuse 1: CE 315mA, 250V, fast type, 5x20mm

Fuse 2: CE 10A, 250V, fast type, 5x20mm

5.2.5. Close the meter case and replace the three screws and rubber jacket.

5.3. To replace the battery:

5.3.1. Turn the meter off and disconnect and leads/equipment from the terminals.

5.3.2. Remove the small screw holding the battery compartment lid, then remove the lid and the old battery.

5.3.3. Place a new 9V (PP3) battery into the meter taking care to observe the correct polarity.

5.3.4. Replace the battery compartment lid and the holding screw.

Range Resolution Accuracy Overload protection

200mV

0.1mV

2V

1mV

20V

10mV

200V

100mV

230V AC

±(0.5%+5)

1000V DC or

750V AC continuous

1000V

1V

±(0.8%+5)

DC Voltage

Range Resolution Accuracy Overload protection

2V

1mV

20V

10mV

200V

100mV

±(0.8%+5)

1000V DC or

750V AC continuous

750V

1V

±(1.0%+4)

AC Voltage

Range Resolution Accuracy Overload protection

200Ω

0.1Ω

2kΩ

1Ω

20kΩ

10Ω

200kΩ

100Ω

600Vp

±(0.8%+5)

2MΩ

1kΩ

Resistance

2MΩ

10kΩ

±(1.5%+5)

Range Resolution Overload protection

1mV

600Vp

Continuity testing

Input impedance: 10MΩ

Input impedance: 10M

Frequency response: 40Hz ~ 400Hz

Original Language Version

5. MAINTENANCE

6. TECHNICAL SPECIFICATIONS

© Jack Sealey Limited

Fig.9

TA200 Issue No: 4(L) - 16/06/14